EXCEL
Program for Computation of Accuracy / Precision Summary Statistics
for Pre-Study Methods Validation
The
attached program was written, validated and made available by P&G
Pharmaceuticals. The program performs computation of Accuracy and
Precision summary statistics as outlined in the manuscript by DeSilva et
al., titled Recommendations for the Bioanalytical Method Validation of
Ligand-Binding Assays to Support Pharmacokinetic Assessments of
Macromolecules,in Pharmaceutical Research [vol 20(11):
1885-1900 (2003)]. The calculated estimates of precision and
accuracy for each Pre-Study validation (QC) sample are obtained using
formulae for an analysis of variance (ANOVA) based on the one-factor
random effects model.
A
few features of this program include computation of a ‘pooled’
estimate for intra-assay imprecision (repeatability), estimation of
‘total random error’ as inter-assay imprecision (intermediate
precision), and use of nominal values for computing the %CV. In
addition to reporting the lower and upper confidence limits for the mean
bias, the summary report includes the 90% Expectation Tolerance Interval
which is useful for predicting future results for In-Study QC
samples. The program is limited to 6 assay runs and 10 replicates
per run.
Please
download the following three files and save to your hard drive:
Installation
and Usage Instructions.pdf (Read this first)
LBA.exe
(Self-extracting files necessary to run the LBA application)
LBA
PDF Files.zip (Contains the
validation related documentation to help modify and improve the
application as needed for individual users.)
For examples of the input and
output files, see the pdf documents included in the Examples
Folder.
Examples Folder:
Batch1.pdf
Batch2.pdf
Batch3.pdf
Batch4.pdf
Batch5.pdf
Batch6.pdf
LBA
application results.pdf
LBA
application sum_tbl.pdf
Disclaimer:
This software was developed and validated at Procter & Gamble
Pharmaceuticals and is being made available free of charge to the
scientific community through AAPS. It is not intended for sale or use
for any commercial purposes. Neither Procter & Gamble
Pharmaceuticals nor AAPS assume any liability for the use of the data
generated using this program. It is the responsibility of the user
to ensure that the software is appropriate for his/her application and to
verify or validate the integrity of the data. Validation data provided in
these files cannot be used by individual users for commercial purposes.
|